Analytical validation of an ultrasensitive multiplexed immunoassay for blood-based biomarkers GFAP, NF-L and tau

The MSD S-PLEX Human Neurology Panel 1 has been validated for measuring GFAP, NF-L, and tau, but minor inter-site variability may affect result consistency.

Read at biorxiv-preprintAdd to compare

Where it sits

this study against the rest of the erythropoietin (epo) corpus
6
Preclinical · this one
26
Observational
1
Open-label
2
Randomised
6
Reviews

Summary and findings

This study validated the MSD® S-PLEX Human Neurology Panel 1 for measuring GFAP, NF-L, and total tau in blood-based samples. The analysis involved multi-site testing to assess precision, accuracy, and stability. No significant differences were found between reagent lots, but minor inter-site variability was noted.

How much of this paper we could read: full text read (0.70). We had a clear abstract, so the summary below closely tracks the paper. What this means →
Intra-plate %CV averaged less than 10%.2026

Abstract

The authors’ words, as biorxiv-preprint supplied them

Abstract Background Blood based biomarkers are increasingly used for non-invasive diagnostics, prognostics, stratification, and therapeutic monitoring, however clinical validation of new biomarkers is often limited by inconsistent assay methods. Assay developers can enable more reliable clinical studies by conducting analytical evaluation to characterize assay performance. In the context of brain injury and neurodegeneration, GFAP, NF-L and tau have emerged as potentially useful biomarkers for stratifying injury severity. Here we demonstrate analytical validation of the MSD® S-PLEX Human Neurology Panel 1, an ultrasensitive multiplex immunoassay for GFAP, NF-L and total tau. Methods Analytical validation testing consisted of multi-site, multi-lot precision, accuracy, stability, dynamic range, and interference. All testing was conducted with plasma, serum, and diluent-based control samples. The precision study was conducted at three sites: Meso Scale Diagnostics, LLC. (Rockville, MD), Johns Hopkins University (Baltimore, MD) and All Children’s Hospital (Tampa, FL). Three kit lots were tested at all three sites for a fully crossed study design. All runs contained an 8-point duplicate calibrator curve, duplicate sample measurements, and control samples. Data were analyzed at each site and then transmitted to MSD for combined analysis. Results There were no significant differences between reagent-lot measurements. Intra-plate %CV averaged less than 10%. Between-site comparison of GFAP, NF-L and tau measurements revealed a small but significant difference in measurements across study sites. Accuracy reported that native samples recover within ± 20% of expected concentration through 20-fold dilution, and spiked samples recover within ± 20% through 256-fold dilution. No substances in a panel of potential interferents affected quantitation by more than ± 20% except for erythropoietin. Kits are shelf stable through 24 months, and GFAP, NF-L and tau in samples can withstand at least five freeze/thaw cycles, storage for 24 hours at 4°C, and storage for four hours at ambient temperature. Conclusions The MSD S-PLEX® Human Neurology Panel 1 has been analytically validated for research use biomarker studies measuring GFAP, NF-L and tau. Characterization of sources of variability have identified minor to moderate inter-site variability, but little to no inter-lot variability. Further work is needed to identify an international reference standard and to evaluate use in specific research contexts.

Background

Not reported in abstract.

Methods

Not reported in abstract.

Results

Not reported in abstract.

Interpretation

Not reported in abstract.

Limitations

Not reported in abstract.

Elsewhere in the Erythropoietin (EPO) corpus

DErythropoietin beyond red blood cells: Systemic and oncological functions (Review).Molecular medicine reports · 2026reviewDSodium-Glucose Cotransporter 2 Inhibitor Use and Levels of Hemoglobin, Markers of Iron Metabolism, and Erythropoiesis: A Systematic Review and Meta-Analysis of Randomized Clinical Trials.JAMA · 2026 · Hemoglobin changed by 0.55 g/dL in the SGLT2 inhibitor group, P<0.001.reviewBResponsiveness to Erythropoiesis‑Stimulating Agents and Hemoglobin Variability in Japanese Patients With End‑Stage Kidney Diseasebiorxiv-preprint · 2026 · ERI was positively correlated with Hb-RSD (r = 0.473, p = 0.003).HumanBBlood Free Carnitine Level is an Independent Determinant of Erythrocyte Membrane Homogeneity and Osmotic Resistance in Maintenance Hemodialysis Patientsbiorxiv-preprint · 2026 · Higher FC levels were associated with increased OFT slope a (β=0.6090, P=0.036).HumanBGenetic mimicry of haematopoietic growth factor signalling: efficacy–safety trade-offs for cytopenias by drug-target Mendelian randomisationbiorxiv-preprint · 2026 · OR 0.65, 95% CI 0.49–0.86, P=0.0026 for lower thrombocytopenia risk with higher MPL expression.HumanCIn vivo HSPC gene therapy of hemoglobinopathies without drug selection of corrected cellsbiorxiv-preprint · 2026 · >70% HbF-positive erythroid cells in treated miceAnimal